NOx-N
Maintenance
3 - 17
ba43110e05
01/2011
Caution
After step 5, you can no longer quit the procedure!
5
Select the
Replace pump tubings
service with
68
and
confirm with
3
. The 1st part of the checklist appears on the
display.
6
Pull the sample tubing out of the overflow vessel and close the
connector.
7
Insert the open end of the tube into a collecting container.
8
Confirm each of the requests of the 1st part of the checklist
with
3
after checking them. After the last confirmation, the
procedure begins immediately and the remaining duration
appears on the display.