NOx-N
Maintenance
3 - 33
ba43110e05
01/2011
3.9
Viewing calibration data of the multi-point calibration
You can view the calibration data of the last multi-point calibration
under this menu item.
Note
The absorbance values of 0.0000 E shown here only serve as wild
cards and are of no significance.
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NO
X
-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
View calibration data
menu item with
68
and
confirm with
3
. The first page of the calibration data
appears.
5
Change to the display of the second page of calibration data
with
3
.
Displaying calibration
data