3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
93
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw in the lower section of the shield cover to remove the cover; to take out the
screw, move slightly upwards for 3mm or so and apply force in a direction perpendicular to the
right clapboard.
Refer to the corresponding section in
3.8.1 Dismantling and replacing the WBC bath
2. Pull out the peripheral fluid tubes attached to the RBC counting bath assembly.
Pay attention to handling the residual fluid; try to prevent it from flowing onto the other
components in the machine, and wipe clean if necessary.
3. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw for grounding on the middle clapboard of the shield case cover and pull out
the grounding wire. See Figure 3-26.
Figure 3-26 Dismantling and replacing the RBC bath assembly (1)
1 - RBC counting chamber
assembly
2 - M3×8 cross-recessed
pan-head combination screw
4. Use 2.5mm hex wrench to remove the two M3×8 stainless-steel inner hex screws affixing the
upper section of the RBC counting bath assembly. See Figure 3-27.