2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
52
Figure 2-40 Digital power input, testing point and indicator light
1 - 5V power inlet
2 - testing point for 5V power supply
3 - 5V power indicator light
⚫
An analog power input of +12V and -12V is enabled by the upper right J10 interface on the main
control panel, as shown in Figure 2-41.
Figure 2-41 Analog power input
1 - ±12V power inlet
2 - +12V testing point
3 - -12V testing point
⚫
A power input of 12V and 24V is enabled by the lower left J17 interface on the main control
panel, as shown in the figure below.