2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
37
Figure 2-30 Checking whether or not the isolation chamber is airtight
5. After inspection, the components and tubing need to be restored to their original locations. Click
the
OK
button in the confirmation box that pops up after the counting bath is drained by opening
the user interface.
If the liquid level in the WBC bath is found to be too high, immediately shut down the device to
prevent any further overflowing.
2.2.12.7 Overflowing of the RBC Bath
The troubleshooting procedure is the same as that in
2.2.12.6 Overflowing of the WBC Bath
; the
only difference is that valve to be inspected is Valve 13. The tubing for inspection is shown in the
picture below, and the isolation chamber to be inspected is the one below the RBC bath.