2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
24
2.2.10.4 Dilution Procedure in Predilute CBC Mode
The dilution procedure in predilute CBC+DIFF mode is shown in Figure 2-12.
Figure 2-12 Dilution Procedure in Predilute CBC Mode
2.2.10.5 Measurement Procedure in Whole-blood CBC+DIFF Mode
In whole-blood mode, a total of 5 time slots for sample measurement are set at 0s~4s, 5s~10s,
11s~22s, 20s~64s, and 64s~72s, respectively, during each of which different steps are taken.
Detailed description:
⚫
Steps taken between 0s~4s (Figure 2-13)
a. Aspirate the whole blood using the sampling syringe.
b. Set the sample probe in the upper position for outer-wall cleaning.
c. Empty WBC/HGB bath.