2 Instrument Configuration
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
67
⚫
Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
⚫
All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
2.4.4 Fine Tuning of the Flow Chamber
Purpose
The fine tuning of the flow chamber can be performed by following the procedures specified in this
section; however, in each case,
Preliminary Steps
need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Preliminary Steps
⚫
Open the right-side door and disassemble the top cover panel.
⚫
Take down the cover panel of the optical system box.
⚫
Turn on the device and launch the software application; then enter the Optical Tuning screen.
Tuning Procedures
1. Use a #2 (Ph2) Phillips screwdriver to loosen the two M3×8 cross-recessed pan-head
combination screws that affix the locking plates of the flow chamber.
Figure 2-55 Fine tuning of the flow chamber
1 - M3×8 cross-recessed pan-head
combination screw (×2)
2 - Adjustment knob of
the flow chamber