6 Alarms and Solutions
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
115
No.
Problem
Name
Problem Description
Solution
12
Abnormal
RBC
counting
RBC clogging.
1. Go to the
Service
>
Maintenance
>
Maintain
screen
and click
Unclog
to start unclogging.
2. Go to the
Service
>
Self-test
>
Valve/Pump Self-test
screen, click
Valve 18
and check if it is working. If it is
OK, please refer to
2.2.12.5 Aperture Clogging of the
3. Go to the
Service
>
Maintenance
>
Maintain
screen
and click
RBC Channel Cleanser Soak
.
13
Waste
overflow
Waste container is
full.
Check the connection of the float sensor at the rear
section of the machine; if the connection is OK, replace
the float sensor for waste overflow detection.
14
Abnormal
reagent
V5 DIL expiration.
Expired reagent or insufficient residual amount. This
means that the reagent has gone out of date or the
amount remaining is insufficient to support hydraulics
operations such as counting. Please follow the
troubleshooting procedures below:
Go to the
Reagent Management
>
Setup
screen, scan
the barcode of the new reagent as shown in the alarm
message, and then load the reagent to resolve the
problem.
Insufficient V5 DIL.
V5 DIL not replaced.
V5 LY2 expiration.
Insufficient V5 LY1.
V5 LY2 not replaced.
V5 LY2 expiration.
Insufficient V5 LY1.
V5 LY2 not replaced.
15
No reagent
No V5 DIL.
1. Check if V5 DIL is running out.
2. If the remaining volume of V5 DIL is a lot, check if the
diluent float sensor is placed correctly. If it does, click
the Remove Error button to automatically resolve the
error.
3. If V5 DIL is running out, replace with new V5 DIL. And
then click the Setup button on the reagent
management interface, and set up the reagent
information according to prompts.
4. Click the Remove Error button to automatically resolve
the error.
5. If the remaining volume of V5 DIL is a lot, or the error
persists after replacing with new V5 DIL, replace the
diluent float sensor.