3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
92
7 - Electrode of the rear bath
8 - O-shaped ring 6.5*1.0
9 - Rear chamber
10 - Optical receiver
11 - Optical filter of the counting
bath assembly
12 - HGB bracket
Installation
⚫
All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
⚫
Pay special attention to the tubing joints and ensure that the connections are sound.
⚫
The aperture, seal gasket, and seal ring need to be correctly positioned and oriented in the right
direction.
Follow the corresponding disassembly steps in reverse order.
Resetting the Machine
1. Reinstall the shield case cover.
2. Close the right side door.
3. Plug the power cord into the back panel of the device and turn on the power switch on the left
panel of the device.
3.8.2 Dismantling and replacing the RBC bath assembly
Purpose
The RBC counting bath assembly can be disassembled and replaced by following the procedures
specified in this section; however, in each case,
3.2 Preparatory Work before Repairs
taken before you begin.
Tools/Spare Parts
⚫
#2 (Ph2) Phillips screwdriver
⚫
2.5mm hex-socket screwdriver
⚫
The replacement WBC counting bath assembly with the same specifications
Preliminary Steps
1. Run the draining sequence of the software to drain any residual liquid inside the machine.
2. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
3. Open the right side door.