3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
77
Figure 3-8 Replacement of display screen (2)
1 - Touch screen
2 - Display screen
3 - Screen support
4 - M3×8 cross recessed pan-head
combination screw (×8)
3.4 Sampling Assembly Replacement
3.4.1 Replacing the Sample probe
Purpose
The sample probe can be disassembled and replaced by following the procedures specified in this
section; however, in each case,
3.2 Preparatory Work before Repairs
begin.
Tools/Spare Parts
⚫
#2 (Ph2) Phillips screwdriver
⚫
Sample probe
Disassembly
1. Pull out the swab's snap ring to detach it from the lower bracket panel. See Figure 3-9.