3 Repairs
Copyright © WOODLEY EQUIPMENT COMPANY LTD.
94
Figure 3-27 Dismantling and replacing the RBC bath assembly (2)
1 - RBC counting chamber
assembly
2- M3×8 stainless-steel inner hex screw
(×2)
5. Carefully and slowly take out the RBC counting bath assembly. Pay attention to the wiring during
the process of moving to avoid breaking the wires.
6. Remove the protective ring from the shield case and open the wire clip to take out the inner wire
along the stretching direction of the wire. Then take out the RBC counting bath assembly as a
whole.
7. The further dismantling of the RBC counting chamber assembly. See Figure 3-28.
Figure 3-28 Dismantling and replacing the RBC bath assembly (4)
1 - Front counting bath
2 - Flat rubber gasket 6*4.5*0.5
3 - RBC aperture
4 - Flat rubber gasket 6*3*0.5
5 - O-shaped ring 5.5*1.0
6 - Electrode of the rear bath
7 - O-shaped ring 6.5*1.0
8 - Rear chamber