Operation,
Screen Layout and Navigation
1-2-3
RGS-2000NG TCAS Test Set Operation Manual
139187 / Rev. 500
Page 12
Subject to Export Control. See Cover Page for details.
September 2021
3.4.3
TCAS MEASUREMENTS SCREEN
The TCAS Measurements Screen allows the user to view the he transmissions of the UUT and to perform a
variety of pulse characteristic, frequency and phase measurements.
NOTE:
DRAGGING THE MOU SE OR FINGER ON THE TOUCH SCREEN OVER THE AXIS AND
GRAPH CAN CHANGE THE HORIZON TAL/VERTICAL SCALES AND
HORIZONTAL/VERTICAL POSITIONS.
3.4.4
MEASUREMENTS - SCOPE MODE
TCAS Scope parameters are displayed by setting the Mode Toggle Switch to Scope Mode. TCAS Scope
parameters can be defined on the Measurements Screen or on the Softkey Menu.
SCREEN SEQUE NCE:
TCAS Screen > Measurements Softkey > Setting: Scope Mode
F i g u re 1 . 2 . 3 - 1 2 T C A S M e a s u re m e n t S cr e e n - S c o p e M o d e
Screen Components
Description
Mode Toggle Switch
The Mode Toggle Switch switches the Measurement Screen mode.
Ch1/Ch2 Settings
Refer to Section
3.4.4.1, Measurements - Scope Mode Channel Softkey
for a description of these settings/controls.
Trigger Settings
Refer to Section
3.4.4.2, Measurements - Scope Mode Trigger Softkey
for a description of these settings/controls.
Horizontal Scale Knob
Selects the Horizontal Scale.
Preset Softkey
Pressing this softkey sets all fields to preset levels and selections.
Trigger Softkey
Accesses Trigger Softkey menu. Refer to Section
Measurements - Scope Mode Trigger Softkey Menu
Channel 1/2 Softkeys
Accesses Channel 1 or Channel 2 Softkey menu. Refer to Section
3.4.4.1, Measurements - Scope Mode Channel Softkey Menu
Ports
Selects the incoming signal Port.
Signal Test
Enables/disables the Signal Test.