Serial Interface Manual
TLS4/TLS-450 Monitoring Systems
Part No. 577013-950, Revision L
422
Function Code:
S51
Version 1
Function Type:
Set LPR Sensor Configuration
Command Format:
Inquire:
Display:
<SOH>SS51QQf
<SOH>IS51QQ
Computer:
<SOH>sS51QQf
<SOH>iS51QQ
Typical Response Message, Display Format:
<SOH>
IS51QQ
JAN 24, 1996 2:54 PM
LINE_PRESSURE_SENSOR CONFIGURATION
DEVICE LABEL CONFIGURED
1 REGULAR UNLEADED ON
<ETX>
Typical Response Message, Computer Format:
<SOH>iS51QQYYMMDDHHmmQQf...
QQf&&CCCC<ETX>
Notes:
1.
YYMMDDHHmm - Current Date and Time
2.
QQ - sensor number (Decimal, 00=All)
3.
f - Configuration flag
0=Off
1=On
4.
&& - Data Termination Flag
5.
CCCC - Message Checksum