4-14
Single Scan Operation
NOTE –
If the profile trace is extremely out of level, change the measurement range to the
maximum range of 2,620kÅ. Level the trace as described above, change to the intermediate
range and repeat the procedure until leveled. The best level is achieved by using the 65kÅ
range.
SOFTWARE LEVELING
Although you have manually adjusted the stage, ensuing scans may show the profile trace slightly
tilted. Software leveling allows the system to quickly level the profile trace without actually having to
completely level the stage. You must software level the stage in order to obtain accurate step height
measurements or accurate readings from analytical functions. Software leveling sets the reference
and measurement cursors at zero to establish a reference for measurements.
Complete the following steps to software level a trace:
1
Position the
R
and
M
cursors along the baseline of the step.
2
Click the
Level
icon, press
F7
, or select
Plot
>
Level
from the system menu bar. The profile trace replots
and levels with the R and M cursor intercepts at zero (see
).
You can also program software leveling into the scan routine to level the trace automatically at the
conclusion of the scan by selecting
Edit
>
Enter Software Leveling
.
Figure 4-14: Cursor Positioning for Software Leveling
Baseline
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......