6-20
Analytical Functions
Figure 6-20: Smoothing Tab of the Data Processing Dialog Box
3
Choose the desired smoothing
Degree
(1, 2, or 3) on the
Smoothing
tab
4
Determine the smoothing band value by performing the
Total Indicated Reading (TIR)
analytical function
on the scan to be smoothed. Follow the instructions in
Activating the Smoothing Function on page 6-
5
In the
Band
field, enter a value that is equal to or greater than the TIR value shown in the
Analytical
Results
area of the
Data Plot
window.
6
Click
OK
to automatically smooth the profile data whenever the current scan routine executes.
7
To clear smoothing, enter zero (0) in the
Band
field of the
Smoothing
tab of the
Data Processing
dialog
box, and then click
OK
.