Electric Force (EFM) Imaging
Electric Field Gradient Detection—Procedures
256
MultiMode SPM Instruction Manual
Rev. B
14.4 Electric Field Gradient Detection—Procedures
Note:
Amplitude detection is the only procedure described here that can be done
without
the Basic Extender Module; however, this method is no longer
recommended (see “Without Basic Extender Module” on page 259).
1. Locate the two toggle switches on the backside of the Basic Extender box (
),
then verify that they are toggled as shown in
Figure 14.4a
Toggle Switches on Back of Basic Extender Module
Table 14.4a
Basic Extender Module Toggle Switch Settings
Note:
The toggle switch combination of Surface Potential = ON and Analog2 = ON is
not recommended
and can produce erratic, unde
fi
ned results.
2. Electrically connect the sample by mounting it to a standard sample disk or stage using
conducting epoxy or silver paint. Ensure the connection is good; a poor connection will
introduce noise.