Setup & Installation
Component List
54
MultiMode SPM Instruction Manual
Rev. B
4. The SPM is connected to the front of the controller with the 37-pin ribbon connector. Verify
that the ribbon cable is securely connected; otherwise, the microscope may not engage or
exhibit other problems.
Note:
If a Basic Extender Module is included, install it now between the SPM and
controller. Verify that it has been properly con
fi
gured per instructions provided
Installation of the Extender Electronics Modules:
of this
manual BEFORE powering the NanoScope controller. If the Quadrex Extender
Module is included, follow the installation instructions provided in the Quadrex
PhaseImaging support note (013-322-000, Rev. C). Connect the Quadrex to
COM1. For the NanoScope IV, refer to that manual for more information.
5. Install the scanner on the support ring with the scanner cap upright. If the motorized scanner
screw will not seat securely into the coupling on the base, manually rotate the screw slightly
until it aligns and seats with the coupling. If installing a non-vertical engagement scanner,
attach retaining springs to the support ring by inserting into holes and securing with screws.
Secure the scanner to the base by inserting the thumb screw (if not a vertical engage
scanner). Plug the scanner into the receptacle on the support ring as shown below.
Figure 3.2e
Installing the Scanner
6. Carefully set the MultiMode’s head atop the SPM scanner. The head should mate
kinematically with the scanner’s contact balls. Secure the head using both retaining springs
by stretching the springs up and hooking them over posts located on both sides of the head.
Plug the head into the receptacle on the support ring as shown below.
Front View
Rear View
Plug scanner into
support ring here.
Motorized scanner
screw must seat
with coupling here.
(Shown: “J” scanner)