Calibration, Maintenance, Troubleshooting and Warranty
Fuse Replacement Procedure
328
MultiMode SPM Instruction Manual
Rev. B
15.15
Fuse Replacement Procedure
The NanoScope III ships with 6 fuses: 3 identical fuses in use and 3 spare fuses. The fuse
characteristics are silk-screened on the back of the NanoScope controller (see
Table 15.15a
NanoScope Controller Fuse Characteristics
To replace a fuse:
1. Remove all power from the system.
2. Remove the lid of the NanoScope controller (4 screws).
3. Using a screwdriver, press the slot in the fuse holder and rotate the fuse a 1/4 turn counter-
clockwise.
4. Gently pull the fuse away from the board.
5. Carefully remove the replacement fuse holder and insert it in the fuse holder.
6. Using a screwdriver, press the slot in the fuse holder and rotate the fuse a 1/4 turn clockwise
until the fuse is secure.
7. Order another spare fuse from Veeco. Be sure to turn off all system power before placing the
spare fuse back in the holder location.
Line
Voltage
Fuse Selection
F1
F2
F3
100V
2.0A
800mA
800mA
120V
2.0A
800mA
800mA
220V
1.0A
400mA
400mA
240V
1.0
400mA
400mA
All Fuses: 250V, Time-lag, 5mm x 20mm