TX6000 Series Operation Manual Ver1.4
117
on every line is set by the pull-up resistance on the HANDLER interface board. The
pull-up resistance is connected to the internal voltage (+5v), or the external
voltage(EXTV:+5v).
The electrical feature of the DC isolation output can be divided into 2 types
shown in Table 4.
Table 4 the electrical feature of the DC isolation output
8.2.3
HANDLER Interface board circuit
The output circuit for the comparison result is shown in the figure 6 below.
the output
signal
the output rated
voltage
the maximum
current
the reference ground
for the circuit
LOW
HIGH
compared
signal
/BIN1
-
/BIN9
/AUX
/OUT
/PHI
/PLO
≤0.5V +5V--+24V
6mA
Internal pull-up
voltage:
TX6000 GND
EXTV1
:
COM1
control signal
/INDEX
/EOM
/ALARM
≤0.5V +5V--+24V
5mA
Internal pull-up
voltage :
TX6000 GND
EXTV2
:
COM2