TX6000 Series Operation Manual Ver1.4
102
The COMParator:BIN:COUNt:DATA? query returns the current comparison result
of the bin count.
Query syntax: COMParator:BIN:COUNt:DATA?
Return format: <BIN1 count>, <BIN2 count>, …, <BIN9 count>, <OUT OF BIN
count>, <AUX BIN count><NL^END>
Where,
<BIN1-9 count> is the count result of BIN1-9, in NR1 data format.
<OUT OF BIN count> is the count result of the OUT OF BIN, in NR1 data format.
<AUX BIN count> is the count result of the auxiliary bin, in NR1 data format.
The COMParator:BIN:COUNt:CLEar command is used to clear all bin count
results.
Command syntax: COMParator:BIN:COUNt:CLEar
For example: WrtCmd (“COMP:BIN:COUN:CLE”)
7.1.15
DCR subsystem commands
The DCR subsystem commands are used for setting and querying the test range,
polarity and level parameters.
Command tree:
DCR
:LEVEL <value>
MIN
MAX
:POL FIX
ALT
:RANGE <value>
:AUTO ON(1)
OFF(0)
The DCR:LEVEL command is used to set the DCR level. The command DCR:LEVEL?
query returns the current level.
(NOTE: Only TX6201 can set different DCR level values, TX6100 and
TX6200 are fixed at 1V.)
Command syntax:
<value>
DCR
:
LEVEL MIN
MAX
Where,
<value> can be NR1, NR2 or NR3 data format.
<MIN> Set the DCR test level to 50mV.
<MAX> Set the DCR test level to 2V.