TX6000 Series Operation Manual Ver1.4
97
The CORRection:USE:DATA? query returns the OPEN/SHORT/LOAD correction
measurement data of all the correction points.
Command syntax: CORRection:USE:DATA?
Return format: <open (n) A>,<open(n) B>,<short (n) A>,<short (n) B>,<load
(n) A>,<load (n) B>
Where,
<open (n) A> is NR3 data format and the primary open correction data at
frequency spot.
<open (n) B> is NR3 data format and the secondary open correction data at
frequency spot.
<short (n) A> is NR3 data format and the primary short correction data at
frequency spot.
<short (N) B> is NR3 data format and the secondary short correction data at
frequency spot.
<load (n) A> is NR3 data format and the primary load correction data at
frequency spot.
<load (n) B> is NR3 data format and the secondary load correction data at
frequency spot.
The CORRection:CLEAR query is used to clear all the correction data.
Query syntax: CORRection:CLEAR
7.1.14
COMParator subsystem commands
The COMParataor subsystem commands are used to set the bin comparator
function including ON/OFF setting, Limit table setting.
Command tree: