TX6000 Series Operation Manual Ver1.4
55
5.5
Operation example of testing capacitance by
multi-frequency list sweep
Test condition:
Function: Cp-D
Level: 1Vrms
Other parameters
Frequency Compare parameter Low limit
High limit
1kHz
Cp(capacitance)
325.0nF
333.0nF
10kHz
D
(
Dissipation
)
0.0001
0.0003
100kHz
D
(
Dissipation
)
0.0060
0.0100
Beep: HIGH LONG
Alarm mode: OUT
Operation steps
1)
Turn on the instrument.
2)
Set basic parameters.
a)
Press [DISP] to enter into the <MEAS DISP> page.
b)
The FUNC zone is currently displayed as Cp-D and the Level zone is
1.000V.
c)
Press [SETUP] to enter into the <MEAS SETUP> page, meanwhile the
following soft keys will be displayed in the soft key zone: MEAS SETUP,
CORRECTION, LIMIT SETUP, SWEEP SETUP and FILE MANAGE.
d)
Press LIST SETUP to enter into the <LIST SWEEP SETUP> page.
e)
Move the cursor to the SWEEP PARAM zone. Then this zone will be
displayed as FREQ[Hz].
f)
Move the cursor to the parameter zone of sweep point 1, then this zone
will displayed as ---.
g)
Press [1], +1 will be displayed in the prompt information zone and the
soft key zone will display available units: Hz, kHz and MHz. Press kHz, this
zone will be changed as 1.00000k.
h)
Press [
] to move the cursor to the LMT zone of sweep point 1, then this
zone will be displayed as ---. The soft key zone will display: LIMIT DATA A,
LIMIT DATA B and OFF.
i)
Press LIMIT DATA A to select the compare primary parameter Cp function.
This zone will be display A and the cursor will automatically move the low
limit zone of sweep point 1.
j)
Press [3][2][5], +325 will be displayed in the prompt information zone
and the following available units will be displayed in the soft key zone: p,
n, μ, m and k. Press n, this zone will be changed as 325.000n. Then the
cursor will automatically move the high limit zone of the sweep point 1.
k)
Press [3] [3] [3], +333 will be displayed in the prompt information zone