parameter, such as Lp-Rd, the step delay time can ensure accurate measurement
while alternate measuring inductive devices by two drive signals.
Operation steps for setting the delay function
Execute the following steps to set the measurement delay time.
1)
Move the cursor to the STEP DLY zone.
2)
Use numeric keys to input delay time. After pressing a numeric key, the
following unit keys will be displayed.
msec
sec
3.5.9
DC resistance range
TX6000 can set the DC resistance range separately. The specific range is the
same as the LCR range, please refer to 3.1.2.
3.5.10
DC level
The DC level of TX6100 and TX6200 is fixed at 1V.
Operation steps for setting the DC level
Execute the following steps to set the measurement delay time.
1)
Move the cursor to the DC LEV zone.
2)
Use numeric keys to input level. After pressing a numeric key, the following
unit keys will be displayed.
mV
V
3)
After pressing the numeric key and choosing the unit, the result will be
automatically converted to the final level.
3.5.11
Deviation test function
The deviation test function can make the deviation value (instead of real test
value) be directly displayed on the screen. The deviation value is equivalent to the
real test value subtracting the pre-set reference value. This function brings great
convenience to observe variations of component parameters with temperature,
frequency, bias. Bias test function can be used for primary or secondary
parameter or primary and secondary parameters meanwhile. The instrument
provides two deviation test modes as below:
ΔABS (Absolute Deviation mode)
The deviation currently displayed is the difference between the test value of
the DUT and the preset reference value. The formula of calculating ΔABS is as
below: