TX6000 Series Operation Manual Ver1.4
109
Control output signal
/INDEX (analog test finished signal), /EOM (test ended and the compared
data effective), /ALARM (the circuit interruption)
Control input signal
/EXT.TRIG (external trigger signal), /Keylock (the key lock).
The signal distribution for the pins is described in Table 2:
Pin
Signal name
Description
1
2
3
4
5
6
7
8
9
10
11
/BIN1
/BIN2
/BIN3
/BIN4
/BIN5
/BIN6
/BIN7
/BIN8
/BIN9
/OUT
/AUX
BIN sorted result
/BIN (BIN number) output are all
open collector output.
12
13
/EXT.TRIG
External trigger: when the trigger
mode is EXT.TRIG, TX6000 will be
triggered by the positive-edge
pulse signal in this pin.
14
15
EXT.DCV2
External DC voltage 2:
The DC provider pin for the
optoelectronic coupling
signal(/EXT_TRIG, /KeyLock,
/ALARM, /INDEX, /EOM)
16
17
18
+5V
The internal power +5V: to use
the internal power is not
recommended. If you use the
internal power, please ensure that
the current is lower than 0.3A and
the signal line is far from the
disturbance source.
19
/PHI
The main parameter is higher: the
test result is greater than the high
limit in BIN1 to BIN9.
20
/PLO
The main parameter is lower: the
test result is less than the low limit
in BIN1 to BIN9.