TX6000 Series Operation Manual Ver1.4
89
FETCh
[:IMP]?
:Source MONitor
:VAC?
:IAC?
:fres :acur?
:bcur?
The FETCh[:IMP]? query directs TX6000 to input the last measurement result to
the output buffer zone.
Query syntax: FETCh[:IMP]?
For example: WrtCmd (“TRIG:SOUR BUS”)
WrtCmd (“TRIG”)
WrtCmd (“FETC?”)
TX6000 applies ASCII to delivery result, details are follows.
On measurement display page, bin NO. display page, bin count display page,
ASCII data output format are described as below:
SN.NNNNNESNN
,
SN.NNNNNESNN
,
SN
,
SN or SNN NL^END
<DATA A>
<DATA B>
<Status>
<BIN number>
Where,
<DATA A>, <DATA B> format: <DATA A> (primary measurement data),
<DATA B> (secondary measurement data)
12-digits ASCII format are as below:
SN.NNNNNESNN
(S:+/-,N: from 0 to 9, E: Exponent Sign)
Status
Description
-1
0
+1
+2
+3
+4
(In data buffer memory) no data
Common measurement data
Analog LCR unbalance
A/D converter is not working.
Signal source is over loading.
Constant voltage cannot be adjusted.
<status> format: When above measurement data is used, <status> data will
display measurement status.
The output format of the <Status> display data uses 2-digits ASCII: SN (S: +/-,
N: from 0 to 4)
NOTE: When <status> is -1, +1 or +2, the measurement data is
9.99999E37. When <status> is 0, +3 or +4, the real measurement data
is beyond the limits.