TX6000 Series Operation Manual Ver1.4
60
Chapter 6
Performance and Test
6.1
Test function
6.1.1
Parameter and symbol
C
:
capacitance
L
:
Inductance
R
:
resistance
Z
:
impedance
Y
:
admittance
X
:
reactance
B
:
susceptance G
:
conductance
D
:
dissipation
θ
:
phase angle Q
:
quality factor
DCR: DC resistance
Test combination
Parameters described above are combined in the following modes:
Primary
parameter
Z, Y
L, C
R
G
Secondary
parameter
θ
(deg phase),
θ
(rad radian)
D, Q, RS, RP, G, Rd X
B
There is no combination for DCR.
Mathematical operation
Operation between the measurement value and the programmable nominal value:
absolute deviation ∆ABS and percent deviation ∆%.
6.1.2
Equivalent mode
Series and parallel
6.1.3
Range
Auto, Manual (Hold, increase and decrease)
6.1.4
Trigger
Internal, external and manual
Internal: Test DUT constantly and display the result
Manual: Press TRIGGER to test once then the result will be displayed.
External: After HANDLER receiving “start” signal, perform a measurement and
output test result.