TX6000 Series Operation Manual Ver1.4
13
RETURN
←
Press the soft key corresponding to your desired parameter. Then press
RETURN
←
to return to upper soft key menu.
9)
Press Y—…
→
, the following parameters will be shown for your choice.
Y-d
Y-r
RETURN
←
Press the soft key corresponding to your desired parameter. Then press
RETURN
←
to return to upper soft key menu.
10)
Press R—…
→
, the following parameters will be shown for your choice.
R-X
Rp-Q
Rs-Q
RETURN
←
Press the soft key corresponding to your desired parameter. Then press
RETURN
←
to return to upper soft key menu.
11)
Press MORE
→
, the following parameters will be shown for your choice.
DCR
MORE
→
3/3
Press DCR, choose the desired parameter. Then press MORE
→
to return to the
first page of soft key menu.
3.1.2
Test range
Measurement range should be selected in accordance with the impedance value
of the tested LCR component.
TX6000 has 10 AC measurement ranges: 3Ω, 10Ω, 30Ω, 100Ω, 300Ω, 1kΩ, 3kΩ,
10kΩ, 30kΩ, 100kΩ.
TX6000 has 11 DCR measurement ranges: 1Ω, 3Ω, 10Ω, 30Ω, 100Ω, 300Ω, 1kΩ,
3kΩ, 10kΩ, 30kΩ, 100kΩ.
Operation steps for setting test range:
1)
Move the cursor to the RANGE zone, the following soft keys will be displayed:
AUTO The soft key is used to set the range mode to AUTO.
HOLD The soft key is used to switch the AUTO mode to the HOLD mode.
When the range mode is set to HOLD, the range will be locked in the
current measurement range. The current measurement range will be
displayed in the range zone.
DECR- The soft key is used to decrease the range under HOLD mode.