| Operation
68
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 6
3. Read the instructions in the newsticker area to learn more about the issue(s) and
available troubleshooting options.
After two cycles of displayed text, the ticker message disappears and is replaced by
the Main Screen shown in Figure 60 above.
4. Tap the red “Health Status Alarm” icon
to display the “Health Status Alarm”
screen shown in Figure 59 and review the alarm list.
5. Correct each issue that has caused an alarm.
As you correct the pending issues, they will disappear from the alarm list on the
“Health Status Alarm” screen one by one.
When the last issue has been cleared, the screen display returns to the healthy state
and the green “Health Status Good” icon
replaces the red
“Health Status Alarm”
icon
.
Alerts
Some events in the biological safety cabinet that do not immediately affect containment or
cleanliness produce alerts. These alerts appear with the visual features shown in the
example in Figure 61.
Figure 61
“Alert” Screen with Filter Replacement Reminder
The “Health Status Alert” icon in the shape of a yellow “Alert triangle” icon appears on a
black background and a short signal sounds. Directly below the “Alert triangle” icon,
descriptive text explains the issue and provides instructions on how to correct it. If there is
more text than will fit on screen, the text scrolls from right to left in “newsticker fashion”.
When multiple alerts occur at the same time, the “Health status” icon has a small blue circle
overlay that shows the number of pending alerts, and additional alert descriptions appear in
the newsticker section.
The alarm volume should be adjusted in the “Alarms and Alerts” screen
according to the sound level of the environment lab.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...