| Operation
46
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 6
Figure 28
Overview of Side-View Symbol Graphics for Unit Status
Three status indicator boxes appear in the center of the Main Screen. They show, from
top to bottom, the state of the window, plus the downflow and inflow statuses.
Status indicator boxes may display with four different status icons:
Working properly, with animated “Downflow” and “Inflow” icons to
show that air barrier has been established
Start-up alert symbol, with animated “Downflow” and “Inflow” icons in
tangerine color to show that air barrier is building up, but not operational
yet
Alarming symbol, with animated “High downflow” and “Low inflow”
icons in red color to warn the user that air barrier is not operational
Alarming symbol, with animated “Low downflow” and “High inflow”
icons in red color to warn the user that air barrier is not operational
Alarming symbol, with animated “Downflow” and “Inflow” icons in red
color to warn the user that air barrier is not operational.
Alarming symbol, alerts the user that window is not ready
The checkmark on the green circle indicates that this particular func-
tion is working properly.
The yellow animated “Timer” icon indicates that the unit is working up
towards an operational state.
The white cross on the red circle indicates that there is a problem with
that particular function.
The light gray dash on the gray circle signifies there is no activity. This
icon indicates, for example, that there is no airflow when the unit is on,
but the fans are off.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...