| Cleaning and Decontamination
128
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 9
Recommended disinfectants:
Predisinfection:
1. Remove all samples from the sample chamber and store them properly.
2. Remove accessories from the safety cabinet and disinfect them using the disinfection
procedure recommended by the manufacturer.
3. The workplate and stainless steel components can be removed from the sample
chamber and disinfected separately.
4. For predisinfection, spray disinfectant on all sample chamber surfaces or wipe the
surfaces using disinfectant.
5. Do not remove the optional UV lamps from the sockets; wipe them thoroughly using a
damp cloth.
6. Switch the device to work mode, move front window to the working position.
7. Allow disinfectant to react as recommended by manufacturer, then operate the safety
cabinet for at least 15 to 20 minutes in the work mode so that released aerosols can
be absorbed by the filters.
Cleaning:
1. Thoroughly remove dirt residues and deposits using a solution of tepid water and
dishwashing agent.
2. Wipe the surfaces clean using a clean cloth and plenty of clear water.
3. Remove the cleaning liquid from the drain pan and wipe all sample chamber surfaces dry.
Final disinfection:
1. Again, spray disinfectant on all sample chamber surfaces or wipe the surfaces clean
with disinfectant.
2. Allow disinfectant to react as recommended by manufacturer.
Compatibility
Chloride-containing disinfectants may damage some surfaces; use only chlo-
ride-free disinfectants or a disinfectant with a low enough chloride content to
have been proved harmless for stainless steel finishes! Disinfectants with an al-
cohol content of more than 70 % may cause embrittlement of plastic compo-
nents after extended exposure. Use only disinfectants with a low alcohol
content. When using a disinfectant with an alcohol content of more than 70 %,
the release limit of 200 g within 2 hours must not be exceeded. Also suited are
disinfectants based on quaternary ammonium compounds.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...