Herasafe 2030i / Maxisafe 2030i
50155215 |
57
Operation |
Chapter 6
Activating and deactivating the internal Gas Supply
The solenoid valve controlling the gas inflow can be opened and closed from the Control Panel.
Proceed as follows to activate the gas inflow:
1. Tap the “Gas” button
.
Gas starts streaming through media valves within the sample chamber, and the
“Gas” button
is displayed with a light gray border.
Proceed as follows to deactivate the gas supply:
2. Tap the “Gas” button
.
The solenoid valve gives a clicking sound, the gas supply to the sample chamber is
turned off, and the light gray border on the “Gas” button
disappears.
Logging into a Passcode-protected Unit
In laboratory environments with exacting biosafety requirements, there may be a need to
restrict access to the biological safety cabinet.
This enhanced level of biosafety is afforded by Secure Mode.
If the unit is in “Secure” Mode, the word “Login” appears in the top right corner of the Info
& Health Status Area, as shown in Figure 42 below.
Figure 42
Info & Health Status Area with “Login” Prompt
As a regular user, your login is not time-limited, that is, the session will never time out.
When you tap “Login”, a login box appears in the center of the screen, then slides up
while a keyboard slides in from the bottom.
NOTE The solenoid valve for activating the internal gas supply is an additional
safety feature that enables the gas supply to the safety cabinet. The main gas
supply is controlled by a media valve or other external shut-off device.
NOTE Secure Mode can only be enabled by an administrator. The process of
enabling Secure Mode is described in the section “Access Control” on
page 112.
Thermo Fisher BSC
11:42 AM | 08/10/2018
Login
NOTE If the unit is set to “Open” mode, the word “Login” will not be displayed
in the top right corner of the screen. Some specialized functions require that
you log in, though. For example, when you tap the “Factory Settings” or “Field
Certification” buttons in the “Settings” screen (see “Settings” on page 73) you
will be prompted to log in.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...