Herasafe 2030i / Maxisafe 2030i
50155215 |
67
Operation |
Chapter 6
Additionally, you will be taken directly to the Health Status Danger screen with its alarm
list shown in Figure 59.
Figure 59
“Health Status Alarm” Screen
Once the ticker message has been visible for two cycles of displayed text, the screen
changes to the appearance shown in Figure 60 below:
Figure 60
Main Screen after two Cycles of Alarm Ticker Message
The “Health status” icon has changed to show the audible alarm is snoozed. To the right
of the “Health status” icon is a countdown timer showing the snooze time remaining, that
is, the time period after which the audible alarm is resumed.
When the issue causing the alarm has been resolved, for example, when the user moves
the front window to the appropriate position, the audible and visual alarms automatically
disappear, and the screen display returns to the healthy state (see “Viewing General
Status Information” on page 70).
Silencing and Handling Alarms
Proceed as follows to silence the audible alarm and handle the problem:
1. When an audible alarm is sounded, go to the biological safety cabinet and tap the
“Snooze” button
to temporarily silence it.
2. Inspect the small blue alarm counter in the red “Health Status Alarm” icon
to
find out whether you have to handle one or more alarms.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...