Herasafe 2030i / Maxisafe 2030i
50155215
|
127
Cleaning and Decontamination
Decontamination Procedure
Several procedures can be applied for decontaminating the safety cabinet. Which
procedure is selected, depends on:
the potential risk imminent in the agents,
the degree of purity required by an experiment or by a work process.
Possible decontamination procedures:
Wipe/spray disinfection:
is the standard disinfection procedure for cabinets used for
microbiological experiments.
UV disinfection:
is particularly suited as an intensifying additional disinfection after a
wipe/spray disinfection.
Sterilization with steam:
can be used for treating the removable stainless steel
components. Examples of autoclavable components are the bezels of the UV lamps, the
workplate or workplate segments and the armrests.
Disinfection with formaldehyde:
can be performed if a sterile sample chamber is
required for the work process. This sterilization procedure is mandatory:
when filters are replaced,
when the device is shut down,
when the device is discarded.
Wipe/Spray Disinfection
The wipe/spray disinfection is performed in three stages:
predisinfection,
cleaning,
final disinfection.
Before cleaning and UV disinfection the covers of the UV lamps must be re-
moved and afterwards installed.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...