| Operation
62
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 6
3. If this happens, tap the “OK” button to close the pop-up window, then use the “Light
Bulb” button
to turn off the sample chamber illumination and/or the “Close
window” button
to close the window.
4. Tap the “UV” button
on the Control panel (again) to start the UV disinfection cycle.
The “UV Disinfection” screen shown in Figure 46 appears. The sample chamber
illumination is off, the fans are on at reduced speed, the gas supply is off, and the
power receptacle is off. The “AutoStop” button
is available as the unit is running.
The “UV”
and “Fan”
buttons are highlighted with a light gray frame to
indicate that both functions are on. The “Fan” button is displayed with a “1/2” label in
a small blue circle to indicate that the fans are running in standby mode.
5. Wait until the UV disinfection cycle is complete.
Completion of the UV disinfection cycle is announced by the alert message shown in
Figure 51.
Figure 51
UV Disinfection Cycle complete Alert
6. Tap the “OK” button to complete the UV disinfection cycle.
Turning the Fan on and off during UV Disinfection
In UV Disinfection mode, the only other button that can be pressed is the “Fan” button.
When the “Fan” button is pressed on, the fans are turned on in Stand-by mode and will
remain on until you turn them off.
Proceed as follows to turn the fan on during a UV disinfection cycle:
1. From within the running UV disinfection cycle, press the “Fan” button
for 5 seconds.
NOTE
If you need to extend or shorten the UV disinfection cycle, you can tap the “Total
cycle time: n hr” label to go the “Disinfection Cycle Settings” screen. The process of
setting cycle time is described in the section “Setting UV Disinfection Cycle Time” on
page 74.
UV disinfection cycle completed.
UV cycle
OK
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...