Table 6 Actions in the Run Monitor screen allowed during a run
(continued)
To…
Do This
Skip to the next step
Select
SKIP
.
Skip works only when the instrument is holding.
Stop the run
Select
STOP
Change the number of cycles for a
cycling stage
Select the
Cycle
field then change the number of cycles (see page
Go to the Main Menu screen
Select (Home). From the
Main Menu
Screen, you can access most of
the instrument functions while the run continues.
View the Status Report
Select
(Status)
.
Changing the number of cycles
You can change the number of cycles for a cycling stage while a run is in progress.
1.
If needed, click the orange
Run Monitor
button to return to the
Run Monitor
screen.
2.
Select the cycles field (below the Stage Name) for a cycling stage, then enter the new value. You
cannot:
• Edit a stage that is already complete.
• Enter a number that is smaller than the number of cycles that have already been completed.
3.
Select
Send
to return to the
Run Monitor
screen.
The run continues, using the new number of cycles.
Pausing a run
1.
In the
Run Monitor
screen, select
Pause Run
.
The
Pause Run
dialog box is displayed:
Chapter 3
Performing a run
Monitoring a run
3
Veriti
™
Thermal Cycler User Guide