6.
Select
(Folder)
to choose a folder in which to save the run method.
• To save the run method so that all users have access to it, select
Shared
.
• To create a new folder for the run method, select
New Folder
at the bottom of the list, enter a
name of up to 12 characters (no spaces), for the new folder, then select
Done
.
• To save the run method to the USB drive, select
USB
(available only when a USB drive is
inserted in the instrument).
7.
Select
Reaction Volume
to enter the reaction volume. (You can change this volume when you
perform a run with this run method.)
8.
Set the cover temperature, if any:
•
Cover Temperature
to enter the temperature for the heated cover.
or
•
Do Not Heat Cover
.
You can change this temperature when you perform a run with this run method.
9.
Select
Notes
, then enter any notes for the method.
10.
Select
Save & Exit
when you are done.
The
Browse Run Methods
screen is displayed.
Editing details
• This section describes how to edit a run method by:
• Adding a stage (“Adding a stage”
)
• Adding a step (“Adding a step”
)
• Editing a temperature or a time (“Editing temperature and time”
• Editing a ramp rate (“Editing ramp rates”
)
• Creating an AutoDelta step (“Creating an AutoDelta step”
• Adding a pause (“Adding a pause”
• Creating a step to program different temperatures on the VeriFlex
™
) (Available only on 96-well instruments.)
Adding a stage
The two types of run method stages are:
•
Cycling –
The steps in a cycling stage are repeated a specified number of times. In general, the
steps in a cycling stage are template denaturation, primer annealing, and primer extension.
•
Holding –
The steps in a holding stage are not repeated.
To add a stage:
Chapter 3
Performing a run
Editing details
3
Veriti
™
Thermal Cycler User Guide
49