4.
Select the check box next to the run method’s original format
(9700 Max
™
Mode
or
9600
Emulation Mode)
then select
(Next).
5.
In the second page, enter the original run method. You can include AutoDelta steps, pauses, and
VeriFlex
™
steps, and you can edit the ramp rates:
Select
(Next), when you are done.
6.
In the Save Run Method page (below), enter a name and select a folder for the run method.
Names are limited to 16 characters. By default, the name begins with “9700-Max
™
-Mode-” or
“9600-Mode-”, depending on the format selected in step 4
Also enter the reaction volume, the temperature for the heated cover (if any), and any notes for this
method.
When you finish, select
Save & Exit.
The Tools Menu is displayed.
Note:
You cannot edit the ramp rates in the method created by the Convert Method wizard. This is
indicated by “9600” or “9700M” in the Time field, when you choose to view ramp rates rather than
time (see Figure 13).
Chapter 4
Utilities
Using GeneAmp
™
PCR System 9700 run methods
4
66
Veriti
™
Thermal Cycler User Guide