Performing a run
Before you begin
Prepare your samples and load the sample block(s) as described in “Loading samples into the
instrument”
IMPORTANT!
You can start a run on multiple instruments at the same time if your instruments are on a
network.
To start a run from the Main Menu screen:
1.
Close the cover.
2.
Select
Browse/New Methods
.
3.
Locate the run method you want to use (see “Viewing a run method”
), then select the
run method.
4.
Select
Start Run
. The
Run Parameters
screen is displayed:
5.
Select
Reaction Volume
to edit the reaction volume. By default, the reaction volume you entered
when you saved the method is displayed.
Reaction volumes are specific to the type of instrument:
• Veriti
™
96-Well
Fast
Thermal Cycler: 5 to 40 µL
• Veriti
™
96-Well Thermal Cycler: 10 to 100 µL
• Veriti
™
384-Well Thermal Cycler: 5 to 30 µL
• Veriti
™
60-Well Thermal Cycler: 10 to 100 µL
6.
Set the cover temperature:
•
Cover Temperature
to enter the temperature for the heated cover.
or
•
Do Not Heat Cover
.
Chapter 3
Performing a run
Performing a run
3
Veriti
™
Thermal Cycler User Guide
57