Getting started
Installing the Veriti
™
Thermal Cycler
Before you start
Site requirements
The instrument is intended for indoor use. Ensure that the installation site:
• Meets the spatial and weight requirements (see page 24)
• Meets environmental requirements (see page 25)
• Is within 1m (3 ft.) of a 800 VA power receptacle
• Is away from water
System specifications
Table 3 Physical dimensions, weight, and power consumption
Parameter
Instrument Footprint
Recommended
Clearance
Total
Length
48.5 cm (19.1 in)
15.5 cm (6.1 in)
[1]
64 cm (15.2 in)
Width
23.7 cm (9.3 in)
0 cm (0 in)
23.7 cm (9.3 in)
Height
[2]
40.0 cm (15.7 in)
15.5 cm (6.1 in)
[3]
55.5 cm (21.8 in)
Weight
10.6 kg (22.6 lbs)
Power Consumption
800 VA – Maximum consumption
∼
260 VA – Average during thermal cycling
∼
10 VA – During sleep/standby
[1]
At the rear of the instrument to ensure adequate airflow and cooling.
[2]
With the heated cover open
[3]
Above the instrument to provide adequate space for opening the heated cover, as shown below.
2
24
Veriti
™
Thermal Cycler User Guide