Installation procedures
Connecting the RS-232 Interface
5-10
Heratherm IMP
Thermo Scientific
—
bb::
is a query, that must be left at „00” for technical reasons;
—
cc
is for a command - specific checksum listed in the table below.
—
<CR>
is for carriage return.
You will receive a response of the following general format:
!:aaaa:bb:XXXXX:cc<CR>
, where:
—
!:
identifies the line as a response to a query;
—
aaaa:
is the parameter address entered with the query;
—
bb:
is the number of payload bytes in hexadecimal code— for example,
1F
for the
decimal value
31
;s
—
XXXXXX:
is the significant status information queried;
—
cc:
is a check sum (technically an inverted XOR of all bytes returned, excluding the
check sum bytes and the <CR> character);
—
<CR>
is for carriage return.
Table 5-1 Terminal Commands for Querying Data
Command Syntax
Response Example
Combined Date and Time
?:0010:00::c1
!:0010:11:
31.07.10
;
01:02:23
:e2
Date
Time
Date only
?:0011:00::c0
!:0011:08:
31.07.10
:d2
Date
Time only
?:0012:00::c3
!:0012:08:
01:02:23
:dc
Time
Temperature Set Value (T1); Current Work Space Temperature (T2); Reference
Temperature (T3); Sample Sensor Temperature (T4)
?:3010:00::c2
!:3010:1f:+125.00;+124.96;+000.000;+000.00:b0
T1
T2
T3
T4
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...