Handling and Control
Settings
7-16
Heratherm IMP
Thermo Scientific
Date and Time
The
Settings -> Time / Date
option allows for switching between the international time and
date display formats and for setting the time and date of the internal clock. There are two
display formats to choose from:
• European date format
DD.MM.YYYY
and 24-hours time format. Example:
07.04.2010
and
15:05
.
• US standard date format
YYYY-MM-DD
and 12-hour time format with
AM/PM
suffix.
Example:
2010-04-07
and
3:05 PM
.
Table 7-19 Entering the Calibration Reference Temperature Manually
Prepare for temperature calibration (see
“Preparing Tem
-
perature Calibration”
and
“Comparison Measurement Pro
-
cedure”
on
page
10-3
).
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to
confirm.
The
Error
menu item from the
Settings
submenu
appears in the multifunctional display pane.
Press
to switch to the
Calibration
menu item and
confirm the selection with
.
In the
Calibration
selection screen, press
to
choose the preselected option
Manual
.
In the settings dialog that appears, set the temperature
measured with the external reference sensor by using
or
and confirm the settings with
.
The newly entered value will be stored and used to
calibrate the internal temperature sensors with the
value measured by the reference sensor.
The display returns to its default mode.
The
Settings
icon in the menu bar will go out.
2010-04-06 1:33
PM
Settings
Error
2010-04-06 1:33
PM
Settings
Calibration
2010-04-06 1:33
PM
Settings
Calibration
Manual
2010-04-06 1:33
PM
Settings
Calibration
36.9°C
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...