Technical Data
13-2
Heratherm IMP
Thermo Scientific
Recovery time (work space
unoccupied), (door opening 30
sec, according to DIN 12880)
typical valvue / max. value
min
5°C (41°F)
7/9
5°C (41°F)
8/10
20°C (68°F)
<1/1
20°C (68°F)
2/3
37°C (99°F)
2/3
37°C (99°F)
2/4
70°C (158°F)
8/10
70°C (158°F)
10/12
Temperature stability, temporal
mid of work space according
to DIN 12880 at 20°C, 25°C,
37°C, (68°F, 77°F, 99°F)
K
≤ ±0,1
≤ ±0,1
Temperature uniformity
according to DIN 12880 at
typical value / max. value
K
5°C (41°F)
≤ ±0.9/≤ ±1.1
5°C (41°F)
≤ ±1.3/≤ ±1.5
20°C (68°F)
≤ ±0.2/≤ ±0.3
20°C (68°F)
≤ ±0.3/≤ ±0.4
37°C (99°F)
≤ ±0.5/≤ ±0.6
37°C (99°F)
≤ ±0.6/≤ ±0.7
50°C (122°F)
≤ ±1.0/≤ ±1.2 50°C (122°F)
≤ ±1.3/≤ ±1.5
70°C (158°F)
≤ ±1.9/≤ ±2.1 70°C (158°F)
≤ ±2.0/≤ ±2.2
BTU output at ambient
temperature of 22°C (72°F)
W
5°C (41°F)
400
5°C (41°F)
1000
20°C (68°F)
100
20°C (68°F)
230
37°C (99°F)
60
37°C (99°F)
170
70°C (158°F)
130
70°C (158°F)
300
Electrical data
Power line voltage (+/- 10%)
100-240
100-240
Power line frequency
Hz
50/60
50/60
Power rating, measured at
ambient temperature of
28°C/82°F
W
450
900
Max. current, measured at
ambient temperature of
28°C/82°F
A
1.9 - 4.5
3.8 - 9.0
Energy consumption,
measured at ambient
temperature of 22°C/72°F
typical value / max. value
5°C/ (41°F)
320/350
5°C/ (41°F)
560/590
20°C (68°F)
80/110
20°C (68°F)
150/180
37°C (99°F)
90/110
37°C (99°F)
170/190
70°C (158°F)
220/240
70°C (158°F)
390/410
Earthing system (e.g. 1/N/PE)
1/N/PE AC
IP protection system
IP 20
Protection class
I
Overvoltage category
according to IEC 60364-4-443
II
Device fusing building site
A
16
Device fusing on PCB
A
2x15
Environmental conditions
Ambient temperature range
°C (°F)
18-28 (64-82)
Table 13-1
Models
IMP 180
IMP 400
920 (230V *) / 700 (120V *)
1360 (230V *) / 1140 (120V *)
4.0 (230V *) / 5.8 (120V *)
6.0 (230V *) / 9.5 (120V *)
* with internal power outlet (optional)
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...