Handling and Control
Programming
7-32
Heratherm IMP
Thermo Scientific
In the list of free program memory slots in the selection
screen that appears (for example:
P5...P10
) press
to choose a target slot for copying, then press
to
confirm the copy target.
In the confirmation dialog for copying that appears,
press
to choose the
Yes
option (default selection is
No)
, then press
to confirm copying.
The display returns to its default mode.
The program you have just copied to the targeted
memory slot may be edited now to adapt it to specific
needs (see the procedure described below under
).
Table 7-34 Editing an Existing Program
Press
to activate the menu bar, then use
or
to select the
Program
icon and press
to
confirm.
In the selection screen shown at left, press
to
switch to the
Edit
option.
In the list of existing programs (for example:
P1...P4
;
default selection is
P1
) shown at left, choose the
program you wish to edit by pressing
, then press
to confirm the selection.
Table 7-33 Copying an Existing Program
2010-04-13 10:25
AM
to
P5
P6 P7 P8 P9 P10
2010-04-13 10:25
AM
Copy
P1 --> P5
Yes No
2010-04-13 10:25
AM
Program
Edit
2010-04-13 10:25
AM
Program
Start
2010-04-13 10:25
AM
Edit
P1 P2 P3 P4
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...