Installation
Installing the Anti-tilt Anchor
3-10
Heratherm IMP
Thermo Scientific
Additionally, the following caution notes must be heeded at all times
:
Risk of overheating with stacked devices
To avoid the risk of electrical components and the outer enclosure overheating
or temperature control failing due to insufficient ventilation, do not exceed the
specified stacking height!
Risk of tipping and dropping of stacked devices
You should be aware at all times that stacked devices do not form a stable unit,
even when the stacking pads and frames are correctly used. The top device may
tip over and drop down when being transported in a stack. To avoid injury to
persons and damage to equipment, do not attempt to move stacked devices as
a unit! Separate and move each device one by one, then re-stack them.
Thermo Scientific don‘t give warranty for stacking equipment of other manufac
-
turer‘s, this occurs at own risk.