Handling and Control
Powering Up
7-4
Heratherm IMP
Thermo Scientific
Powering Up
1. Plug the power plug of the refrigerated incubator into a suitable protection-earthed AC
power outlet.
In the display window on the front panel the readiness indicator icon (rightmost icon in
the status display area at D4 in
) is illuminated.
2. Keep the
On/Off
button depressed for two seconds.
An initialization routine will be run after the refrigerated incubator has been powered up.
On completion of the initialization, the display will light up and the current work space
temperature will appear in the temperature display field (item D1 in
). The refrigerated incubator is ready for use now.
Timer
Allows for having the refrigerated incubator turn on or off upon expiry of a
user-specified countdown period or at a fixed on or off time, or having it
operate on a complete weekly schedule of daily on and off times. When
the user enables an “on timer” the display goes out. A rotating hand in
the
Timer
icon and the illuminated readiness indicator icon in the status
display area indicates that the timer is running.
Instructions:
Power Outlet (optional)
Toggles the AC outlet in the work space of the refrigerated incubator on
and off.
Instructions:
.
Fan
Shows function status of the fan.
Settings
Invokes a submenu with the following functions:
- Read access to error log
- Calibrating the refrigerated incubator
- Setting date and time
- Toggling the temperature display unit between °C and °F
- Selecting the timer mode of operation (countdown / fixed time of day /
weekday timer)
- Entering a configuration control code
- Setting over temperature cut-out according to DIN 12880 (TWB - Tem
-
perature selection limiter)
(Instructions:
Table 7-3 Menu Bar Icons
Icon
Function
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...