Installation
Installing the Anti-tilt Anchor
Thermo Scientific
Heratherm IMP
3-9
Floor stand refrigerated incubators
Remove the screws.
Attach the end of the retaining bracket that is facing downward to the unit.
Align the device at roughly 90°, +/-20° to the retaining bracket.
Affix the retaining bracket to the wall.
Heratherm floor stand refrigerated incubators must always be attached to the
wall using two (2) retaining brackets on the outer left and right side on the
back of the unit.
When operating under maximum environment- and condition of employment
and when using low setpoint temperature can lead to performance losses in
stacked devices.
Unsafe part of the building!
Install the anti-tilt anchor to a solid part of the building, which is able for shoring
loads.
The installation has to be carried out by qualified personnel only.
The connection to the building must be carried out with appropriate screws and
dowels according to the consistence of the building part.
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...