Thermo Scientific
Heratherm IMP
13-1
13
Technical Data
The technical data are valid only for an empty device equipped with three shelves and a spray-painted outer
enclosure. Options may have an impact on the specified performance.
Table 13-1
Models
IMP 180
IMP 400
Parameter
Unit
Chamber volume
L/cu ft
178/6.3
381/13.5
Chamber temperature range
°C/°F
5-70 (41-158)
Unit dimensions
Internal dimensions
W x D x H
mm/in
494x590x708 /
19.5x23.2x27.9
600x590x1335 /
23.6x23.2x52.6
Internal dimensions (useable
space) W x D x H
mm/in
464x543x708 /
18.3x21.4x27.9
544x524x1335 /
21.4x20.6x52.6
External dimensions (without
feet / castors W x D x H
mm/in
640x738x920 /
25.2x29.1x36.2
778x770x1545 /
30.6x30.3x60.8
Footprint
m2/ft2
0.47 / 5.1
0.56 /6.0
Number of shelves:
standard/maximum
2 / 9
2 / 17
Number of shelf positions
19
39
Shelf size
mm/in
439x505 / 17.28x19.88
528x498 / 20.79x21.06
Loading capacity per shelf
kg/lbs
25/55
30/66
Loading capacity of unit
kg/lbs
75/165
75/165
Weight of the unit
kg/lbs
84/185
167/368
Weight of unit incl. packing
kg/lbs
97/214
216/476
External dimension incl.
packing W x D x H
mm/in
960x770x1200 /
37.8x30.3x47.2
1030x1100x1900 /
40.6x43.3x74.8
Access port diameter
mm/in
42 / 1.7
42 / 1.7
Temperature Performance
Heat up / cool down time (work
space unoccupied), from
ambient temperature 22 °C
(72 °F) to
(according to DIN 12880)
min
5°C/ (41°F)
55/60
5°C/ (41°F)
55/60
20°C (68°F)
7/8
20°C (68°F)
9/10
37°C (99°F)
31/34
37°C (99°F)
52/56
70°C (158°F
61/66
70°C (158°F
92/101
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...