Handling and Control
Settings
Thermo Scientific
Heratherm IMP
7-15
Calibration
The
Settings -> Calibration
menu item enables the user to initiate a temperature calibration
process (see
“Temperature Calibration Procedure”
on
page 10-4
) for the built-in temperature
sensors and choose whether calibration should be accomplished manually or automatically:
• The
Manual
option allows for entering an absolute temperature directly, as measured - for
example, using an external reference sensor.
Table 7-18 Reading the Error Log
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to
confirm.
Press
to select the
Error
item from the
Settings
submenu.
The first entry of the error log is displayed, named
“
Error 0
.”
Press
to go to the next entry (or
to go back to
the previous one).
After the entry numbered
21
the display wraps and
returns to the beginning of the error log, displaying
Error
0
again.
To exit from the error log and return to normal display
mode press
twice.
The
Settings
icon in the menu bar will go out.
2010-04-06 1:33
PM
Settings
Error
2010-04-06 1:36
PM
Error 0
2010-04-06 1:31
PM
Fan Error
(E009)
2010-04-06 1:37
PM
Error 1
2010-04-06 1:34
PM
Heat Relay
(E109)
Calibration Prerequisites
Maintain the ambient conditions within the specified limits of the
refrigerated incubator before launching calibration.
Varying ambient conditions may impact the result of the calibration
routine, which may lead to mis-adjustment of the controller and
unreliable temperature control operation.
In individual cases, the display may differ from the actual temperature
value and can be corrected by using the function calibration.
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...