TWB Setting/Activating
With menu point
Settings -> Over Temp
the TWB (adjustable temperature selection limiter)
of the refrigerated incubator can be set, according to DIN 12880 class 2.
Tabelle 7-24 Setting the over temperature cut-out
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to
confirm.
Press
to invoke menu point
Over Temp Limit
of the
sub menu
Settings
.
Press
to change to
on
and press
to confirm.
Press
or
to select the Set Temperatur, e.g.
39 °C
.
Press
to confirm the selection.
Pressing
the display returns to the standard
display with activated
Over Temp
.
The
Settings
icon in the menu bar will go out.
Set temperature of the Over Temp
In the unlikely event that the temperature in the chamber
surpasses the set temperature of the Over Temp, the unit will
switch off the Peltier element. The operator needs to re-activate
the unit by pressing
Summary of Contents for IMP 180
Page 1: ......
Page 6: ...Contents iv Heratherm IMP Thermo Scientific...
Page 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Page 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Page 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Page 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Page 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Page 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Page 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Page 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...