Dionex ICS-5000
+
Installation Instructions
44
Doc. 065447-02 12/14
3. Connect one end of the waste line to port 6 on the injection valve. For a
capillary system with a two-piece waste line, connect the black tubing to
the injection valve.
NOTE
In a capillary system, port 6 is labeled
SAMPLE IN
. In
an analytical system, port 6 is labeled
W
.
4. Route the line to the Dionex AS-AP waste port and insert the line into one
of the round openings in the waste port (see
Figure 3-15. Injection Valve Waste Line Installed in
Dionex AS-AP Waste Port
Waste line from
injection valve
Waste line from
syringe
Note: The injection valve
waste line is green in
analytical systems and
black in most capillary
systems.